Replace a Broken Liner
11. Unscrew the nut that retains the analytical
column.
12. Remove the analytical column with its ferrule.
13. Unscrew the retaining nut at the bottom of the
injector.
14. Remove the terminal fitting with its silver seal.
CAUTION
Glass splinters from the broken
liner will fall from the injector.
15. With the aid of a pipe cleaner, remove the
possible glass fragments from the vaporization
chamber.