Replace a Broken Liner
20. Holding the new liner with tweezers, place a
graphite seal over the liner, making sure to
leave a distance of about 10 mm between the
seal and the liner end.
21. Using tweezers, insert the liner into the injector
and push it gently towards the bottom fitting.
Make sure you insert the bevelled end of the
liner towards the bottom of the injector.
CAUTION
Be careful not to damage the
graphite seal or allow graphite
to entering the liner. Should this
occur, clean the liner with an
inert gas.
22. Tighten the liner cap using the liner cap wrench
provided with the GC.
Maintenance and Troubleshooting Manual
Maintaining a Split/Splitless Injector
Chapter 2
45