Replace a Broken Liner
14. Use tweezers to reinsert the spacer inside the
injector, complying with the correct orientation.
15. Holding a liner with tweezers, place a graphite
seal over it, making sure to leave a distance of
about 10 mm between the seal and the liner end.
16. Using tweezers, insert the liner into the injector
and push it gently towards the bottom.
CAUTION
Be careful not to damage the
graphite seal or allow graphite
to entering the liner. Should this
occur, clean the liner with an
inert gas.
PRECAUTIONS
Maintenance and Troubleshooting Manual
Chapter 9
Maintaining a PTV Injector
161