Adjust A Quartz Wool Packing - Thermo Scientific TRACE GC Ultra Maintenance And Troubleshooting Manual

Hide thumbs Also See for TRACE GC Ultra:
Table of Contents

Advertisement

Adjust a Quartz Wool Packing

OPERATING SEQUENCE
Adjust a Quartz Wool Packing
Materials needed
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the injector must be cooled to room temperature.
PRECAUTIONS
1. While in stand-by condition, press OVEN to
access the oven control table.
and press OFF.
2. Scroll to
Temp
3. Press RIGHT INLET.
and press OFF.
4. Scroll to
Temp
Maintenance and Troubleshooting Manual
Screwdriver (P/N 205 026 03)
Graphite seal (P/N 290 134 17)
Non metallic sharp tool
Thin (<2 mm diameter) non metallic bar (e.g. 1.6 mm PEEK tube)
Septum (P/N 313 132 25)
Tweezers
OVEN
Temp
Initial time
Ramp 1
RIGHT INLET (PTV)
Temp
Pressure
Mode
Chapter 9
Maintaining a PTV Injector
.
50
50 <
5
10.0
50
50 <
100
100
split
165