Chapter 17
Maintaining Detectors in Stack Configuration
10. Disconnect the signal, excitation and heater
cables from the ECD detector
11. Remove the heat shield.
12. Remove the heater from the top of the ECD
detector unscrewing and removing the 4 mm
Allen fixing screw. Use a 3 mm Allen wrench.
352
Clean or Replace the Collecting Electrode (FID on ECD)
Maintenance and Troubleshooting Manual