Thermo Scientific TRACE GC Ultra Maintenance And Troubleshooting Manual page 294

Hide thumbs Also See for TRACE GC Ultra:
Table of Contents

Advertisement

Chapter 14
Maintaining a Flame Photometric Detector
6. Using the tool provided with the GC, loosen the
fixing nut on the base of the detector and
remove it.
CAUTION
Pay attention not to lose the
aluminium ring inserted
between the detector head and
the base body.
7. Using the 5 mm wrench provided with the GC,
loosen the worn jet and remove it from the
detector base body.
294
Maintenance and Troubleshooting Manual
Replace the Jet

Advertisement

Table of Contents
loading

Table of Contents