Clean the Jet

OPERATING SEQUENCE
Clean the Jet
Materials needed:
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
Maintenance and Troubleshooting Manual
Detector fixing tool (P/N 205 021 50)
5 mm wrench
Forceps or tweezers
Ultrasonic cleaning bath
Liquid detergent
Clean compressed air
GC-grade methanol
Paper towels
Maintaining a Flame Photometric Detector
Reagents Safety Information
Chapter 14
for more information.
273