Replace the UV Lamp
4. Remove the lamp with its Viton
the lamp holder and store it.
CAUTION
Avoid touching the lamp and
particularly the lamp window
and the O-ring with your fingers.
Use cotton gloves during this
operation.
5. Place the proper Viton
lamp so to contact its flange.
6. Insert the lamp into the holder, pushing it in
until it stops.
CAUTION
While inserting the lamp, pay
attention not to damage the
contact pin inside the holder.
Maintenance and Troubleshooting Manual
TM
O-ring from
TM
O-ring on the new
Maintaining a Photoionization Detector
Chapter 15
321