Chapter 15
Maintaining a Photoionization Detector
9. Rinse the lamp window with GC-grade
methanol or acetone.
10. Wipe the window using a clean paper towel.
11. Air dry the lamp.
TM
12. Insert the Viton
it in until it touches the flange.
13. Insert the lamp into the holder, pushing it in
until it stops.
CAUTION
While inserting the lamp, pay
attention not to damage the
contact pin inside the holder.
326
O-ring on the lamp, pushing
Clean the Lamp Window
Maintenance and Troubleshooting Manual