Chapter 9
Maintaining a PTV Injector
18. Place the septum support into its housing inside
the injector.
19. Reinsert the septum holder with the septum.
20. Tighten the injector cap to finger tight.
CAUTION
Do not overtighten the injector
cap. You could damage the
septum and affect performance.
21. Press RIGHT CARRIER, scroll to
press ON.
22. Press OVEN, scroll to
23. Press RIGHT INLET, scroll to
24. Perform a leak check according to
Leak Check (Capillary Column)
Sequence on page 181.
156
Pressure
and press ON.
Temp
and press ON.
Temp
Performing a
Operating
and
Maintenance and Troubleshooting Manual
Clean or Replace the Liner