Chapter 3
Maintaining a Large Volume Splitless Injector
9. Insert a new septum into the septum holder.
CAUTION
Use tweezers to avoid touching
the septum with your fingers.
10. Tighten the injector cap to finger tight.
CAUTION
Do not overtighten the injector
cap. You could damage the
septum and affect performance.
11. Press LEFT CARRIER or RIGHT CARRIER,
depending on which injector is operating, scroll
and press ON.
to
Pressure
12. Press OVEN, scroll to
13. Press LEFT INLET or RIGHT INLET, depending on
which injector is operating, scroll to
press ON.
14. Perform a leak check according to
Leak Check (Capillary Column)
Sequence on page 181.
60
and press ON.
Temp
Temp
Performing a
Operating
and
Maintenance and Troubleshooting Manual
Replace the Septum