Chapter 17
Maintaining Detectors in Stack Configuration
21. Reconnect the signal and ignition cables to the
NPD and the signal, excitation and heater
cables to the ECD detector.
22. Press LEFT DETECTOR or RIGHT DETECTOR,
depending on the location of the ECD, and reset
the ECD to the required operating conditions.
23. Press AUX.
24. Scroll to
Detector
25. In the
AUX DETECTOR (NPD)
auxiliary NPD detector to the required
operating conditions.
362
and press ENTER.
table, reset the
Clean or Replace the Collecting Electrode (NPD on ECD)
Maintenance and Troubleshooting Manual