Chapter 13
Maintaining a Nitrogen Phosphorus Detector
10. Place the jet into its housing and tighten it with
the proper tool.
11. Place the detector cell on the detector base body
and reconnect the fixing screw on the front of
the detector cell.
12. Reconnect the signal and ignition cables to the
detector.
13. Reset the detector to the required operating
conditions. Refer to the relevant chapter of the
TM
TRACE
GC Ultra Operating manual for
more information.
246
Maintenance and Troubleshooting Manual
Clean the Jet