Clean the Interferential Filter
9. Insert the cleansed filter into its housing. The
mirror face must be oriented towards the flame.
10. Reassembly the photomultiplier tube and the
detector body, then fix them together tightening
the knurled nut.
11. Reconnect the signal, excitation voltage and
ignition/heating cables to the detector.
12. Switch on the GC power supply and set the
detector to the required operating conditions.
Refer to the TRACE
manual for more information.