Chapter 2
Maintaining a Split/Splitless Injector
15. Press LEFT INLET or RIGHT INLET, depending on
which injector is operating, scroll to
press ON.
16. Perform a leak check according to
Leak Check (Capillary Column)
Sequence on page 181.If leaks are detected,
check the critical connections with an electronic
leak detector to locate the possible leaks.
17. When the injector is free of leaks, reinstall the
GC top cover.
18. Set the normal injector and GC working
condition.
52
and
Temp
Performing a
Operating
Replace the Split Line Tubing
Maintenance and Troubleshooting Manual