Clean the Mirror Metal Plug
9. Insert the mirror plug in the detector body and
fix it with the relevant Allen screws.
NOTE
Before reinserting the plug, inspect
the graphite seal inside the plug's
housing. If the seal is damaged
and could not ensure tightness, it
must be removed and replaced
with a new one.
10. Place the detector on its base body, paying
attention that the aluminium ring has been
inserted in the correct position, then tighten the
fixing nut.
Maintenance and Troubleshooting Manual
Maintaining a Flame Photometric Detector
Chapter 14
285