Chapter 3
Maintaining a Large Volume Splitless Injector
13. Unscrew the retaining nut at the bottom of the
injector.
14. Remove the terminal fitting with its silver seal.
CAUTION
Glass splinters from the broken
liner will fall from the injector.
15. With the aid of a pipe cleaner, remove the
possible glass fragments from the vaporization
chamber.
16. Reinsert the silver seal and the terminal fitting.
17. Tighten the nut that retains the terminal fitting.
70
Replace a Broken Liner
Maintenance and Troubleshooting Manual