Thermo Scientific TRACE GC Ultra Maintenance And Troubleshooting Manual page 370

Hide thumbs Also See for TRACE GC Ultra:
Table of Contents

Advertisement

Chapter 17
Maintaining Detectors in Stack Configuration
3.
In the
AUX DETECTOR (FID)
and press OFF.
Flame
4. Scroll subsequently to
and press OFF.
5. Press LEFT DETECTOR or RIGHT DETECTOR,
depending on the location of the ECD, to access
the ECD control table.
6. Scroll subsequently to
and press OFF.
temp
7. Scroll to
Mkup(N2)
370
table, scroll to
and
Base temp,H2
and
Base temp
ECD
and press OFF.
AUX DETECTOR (FID)
Flame
Air
Base temp
Signal pA
Ign. tresh
Flameout retry
H2
Air
XXX DETECTOR (ECD)
Base temp
ECD temp
Ref current nA
Freq kHz
Pulse amp V
Pulse width us
Mkup (N2)
Maintenance and Troubleshooting Manual
Replace the Silver Seal (FID on ECD)
.
Off <
300
Off
(5.4)
2.0
Off
35
Off
350
Off
.
250
Off <
250
Off
1.0
(2.15)
1.0
40
Off
25

Advertisement

Table of Contents
loading

Table of Contents