Clean or Replace the Collecting Electrode (FPD on ECD)
3. Using the tool provided with the GC, loosen the
fixing nut on the base of the FPD detector and
remove it.
CAUTION
Pay attention not to lose the
aluminium ring inserted
between the detector head and
the base body.
4. Disconnect the signal, excitation and heater
cables from the ECD detector.
Maintenance and Troubleshooting Manual
Maintaining Detectors in Stack Configuration
Chapter 17
365