Chapter 9
Maintaining a PTV Injector
5. When the inlet reaches the room temperature,
press RIGHT CARRIER.
6. Scroll to
Pressure
7. Unscrew the injector cap.
8. Remove the septum holder with septum, then
the septum support.
166
and press OFF.
RIGHT CARRIER (PTV)
Col. flow
Pressure
Flow mode
Maintenance and Troubleshooting Manual
Adjust a Quartz Wool Packing
2.5
<
100
100
con pres