Chapter 13
Maintaining a Nitrogen Phosphorus Detector
13. Place the collecting electrode into its housing.
Ensure the insert hole is perfectly vertically
aligned to the hole present on the cell.
14. Using forceps, reconnect the gold contact pin.
15. Remount the connector holder assembly
verifying the contact efficiency.
254
Clean the Collecting Electrode
Maintenance and Troubleshooting Manual