Replace the Interferential Filter

OPERATING SEQUENCE
Replace the Interferential Filter
Materials needed
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS
Maintenance and Troubleshooting Manual
Detector fixing tool (P/N 205 021 50)
Interferential filter (see Table 14-2 for available filters)
Paper towels
Table 14-2. Available interferential filters for the FPD
Target compound
Phosphorus
Sulphur
Organotin
Maintaining a Flame Photometric Detector
detector
P/N
281 071 00
281 070 00
281 070 01
Chapter 14