Clean the Collecting Electrode
Press LEFT DETECTOR or RIGHT DETECTOR,
1.
depending on the location of the FID detector,
to access the FID control table. In the following
example, a FID installed on the right channel is
considered.
and press OFF.
2. Scroll to
Flame
3. Scroll subsequently to
and press OFF.
4. Disconnect the signal and ignition cables from
the detector.
5. Loosen the fixing nut on the front of the
detector cell and remove it.
6. Unscrew and remove the cap of the detector
cell.
Maintenance and Troubleshooting Manual
and
Base temp,H2
Air
RIGHT DETECTOR (FID)
Flame
Base temp
Signal pA
Ign. tresh
Flameout retry
H2
Air
Mkup N2
Chapter 11
Maintaining a FID
.
Off <
300
300
(5.4)
2.0
Off
35
Off
350
Off
30
30
203