Chapter 2
Maintaining a Split/Splitless Injector
OPERATING SEQUENCE
Clean or Replace the Liner
Materials needed
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the injector must be cooled to room temperature.
PRECAUTIONS
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
1. While in stand-by condition, press OVEN to
access the oven control table.
and press OFF.
2. Scroll to
Temp
34
Liner cap removal tool (P/N 205 070 10)
Glass liner (see Table 2-2, on page 29)
Graphite seal (P/N 290 334 06)
Ultrasonic bath
Methanol/acetone mixture (1:1)
Non metallic sharp tool
Septum (P/N 313 032 11)
Tweezers
Reagents Safety Information
Temp
Initial time
Ramp 1
Maintenance and Troubleshooting Manual
Clean or Replace the Liner
for more information.
.
OVEN
50
50 <
2
10.0