Replace the Ignition Assembly
7. Remount the new connector holder assembly
and fix it to the detector cell using the two
screws. Make sure the connecting fork is placed
downwards to achieve the correct contact with
the jet.
8. Place the detector cell on the detector base body
and tighten the retaining screw.
9. Reconnect the signal and ignition cables to the
detector.
10. Reconnect the signal and ignition cables to the
detector.
11. Reset the detector to the required operating
conditions. Refer to the relevant chapter of the
TM
TRACE
GC Ultra Operating manual for
more information.
Maintenance and Troubleshooting Manual
Chapter 11
Maintaining a FID
213