3. Press LEFT INLET or RIGHT INLET depending on
which injector is operating. In the following
example, a S/SL injector installed on the right
channel is considered.
and press OFF.
4. Scroll to
Temp
5. When the inlet reaches the room temperature,
press
LEFT CARRIER
depending on which injector is operating.
6. Scroll to
Pressure
7. Unscrew the injector cap.
8. Remove the septum holder with septum, then
the septum support.
Maintenance and Troubleshooting Manual
or
RIGHT CARRIER
and press
.
OFF
Maintaining a Split/Splitless Injector
RIGHT INLET (S/SL)
Temp
Pressure
Mode
RIGHT CARRIER (S/SL)
Pressure
Col. flow
Lin. Veloc.
Chapter 2
250
250 <
100
100
split
30.0
30.0
3.00
<
(60.9)
35