Chapter 13
Maintaining a Nitrogen Phosphorus Detector
OPERATING SEQUENCE

Replace the NPD Thermionic Source

Materials needed:
CAUTION
Thermionic sources are fragile. Handle them with care. Gently remove the plastic
tube protecting the source ceramic during shipment.
Especially avoid any excessive transverse force to the source ceramic.
Before inserting the new source into the holder check that the PTFE insulator is
properly interposed.
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS
260
Screwdriver
1 mm Allen wrench
Thermionic Source TS-1 (P/N 465 002 50) or TS-2 (P/N 465 002 55)
Replace the NPD Thermionic Source
Maintenance and Troubleshooting Manual