Chapter 12
Maintaining an Electron Capture Detector
8. Remount the electrode without touching it. Use
tweezers.
9. Remount the detector cover with the chimney.
10. Reconnect the signal and ignition cables to the
detector.
11. Reset the detector to the required operating
conditions.
228
Clean the Collecting Electrode
Maintenance and Troubleshooting Manual