Chapter 8
Maintaining a Purged Packed Column Injector
OPERATING SEQUENCE

Clean or Replace the Liner

Materials needed
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the injector must be cooled to room temperature.
PRECAUTIONS
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
1. While in stand-by condition, press OVEN to
access the oven control table.
and press OFF.
2. Scroll to
Temp
136
Glass liner (see item 8, Table 5-1)
Ultrasonic bath
Methanol/acetone mixture (1:1)
Tweezers
Reagents Safety Information
Temp
Initial time
Ramp 1
Maintenance and Troubleshooting Manual
Clean or Replace the Liner
for more information.
.
OVEN
80
80 <
200
Off