Thermo Scientific TRACE GC Ultra Maintenance And Troubleshooting Manual page 351

Hide thumbs Also See for TRACE GC Ultra:
Table of Contents

Advertisement

Clean or Replace the Collecting Electrode (FID on ECD)
8. Wait until the detectors temperature is near to
ambient, then disconnect the signal and ignition
polarization cables from the FID.
9. Loosen the fixing screw on the front of the FID
cell and remove it.
Maintenance and Troubleshooting Manual
Maintaining Detectors in Stack Configuration
Chapter 17
351