Thermo Scientific TRACE GC Ultra Maintenance And Troubleshooting Manual page 109

Hide thumbs Also See for TRACE GC Ultra:
Table of Contents

Advertisement

Replace the Rotary Valve Seal
28. Press OVEN, scroll to
29. Perform a leak check according to
Leak Check (Capillary Column)
Sequence on page 181.
30.
From the data system in use, enable the
automatic actuator to open and close the rotary
valve of the OC injector by using the direct
command
OC Actuator
the OC valve lever as shown in the figure on the
right.
Maintenance and Troubleshooting Manual
and press ON.
Temp
Performing a
Operating
. The actuator move
Maintaining a Cold On-Column Injector Direct Narrow-Bore
Open
Close
Chapter 5
109

Advertisement

Table of Contents
loading

Table of Contents