Chapter 14
Maintaining a Flame Photometric Detector
10. Insert the mirror plug in the detector body and
fix it with the relevant Allen screws.
NOTE
Before reinserting the plug,
inspect the graphite seal inside the
plug's housing. If the seal is
damaged and could not ensure
tightness, it must be removed and
replaced with a new one.
11. Place the detector on its base body, paying
attention that the aluminium ring has been
inserted in the correct position, then tighten the
fixing nut.
290
Clean the Flame-side Heat Shield
Maintenance and Troubleshooting Manual