Chapter 9
Maintaining a PTV Injector
3. Press RIGHT INLET.
and press OFF.
4. Scroll to
Temp
5. When the inlet reaches the room temperature,
press RIGHT CARRIER depending on which
injector is operating.
6. Scroll to
Pressure
7. Unscrew the injector cap.
8. Remove the septum holder with septum, then
the septum support.
152
and press OFF.
RIGHT INLET (PTV)
Temp
Pressure
Mode
RIGHT CARRIER (PTV)
Col. flow
Pressure
Flow mode
Maintenance and Troubleshooting Manual
Clean or Replace the Liner
50
50 <
100
100
split
2.5
<
100
100
con pres