Clean or Replace the Liner
14. Tighten the liner cap using the liner cap wrench
provided with the GC.
15. Place the septum holder with the septum onto
the inlet port, then screw and tighten the ring
nut up to the stop.
16. Tighten the injector cap to finger tight.
CAUTION
Do not overtighten the injector
cap. You could damage the
septum and affect performance.
Maintenance and Troubleshooting Manual
Maintaining a Large Volume Splitless Injector
Chapter 3