Thermo Scientific TRACE GC Ultra Maintenance And Troubleshooting Manual page 355

Hide thumbs Also See for TRACE GC Ultra:
Table of Contents

Advertisement

Clean or Replace the Collecting Electrode (FID on ECD)
21. Reconnect the signal and ignition cables to the
FID and the signal, excitation and heater cables
to the ECD detector.
22. Press LEFT DETECTOR or RIGHT DETECTOR,
depending on the location of the ECD, and reset
the ECD to the required operating conditions.
23. Press AUX.
24. Scroll to
Detector
25. In the
AUX DETECTOR (FID)
auxiliary FID detector to the required operating
conditions.
Maintenance and Troubleshooting Manual
and press ENTER.
table, reset the
Maintaining Detectors in Stack Configuration
Chapter 17
355