Chapter 14
Maintaining a Flame Photometric Detector
OPERATING SEQUENCE
Clean the Mirror Metal Plug
Materials needed
•
•
•
•
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
282
Detector fixing tool (P/N 205 021 50)
1 mm Allen wrench
Paper towels
Methylene chloride or methanol
Reagents Safety Information
Maintenance and Troubleshooting Manual
Clean the Mirror Metal Plug
for more information.