Replace the Jet
7. Place a new jet into the housing and tighten it
with the proper tool. Ensure the jet is perfectly
vertically aligned to avoid damaging its ceramic
part.
8. Place the detector cell on the detector base body
and tighten the fixing screw on the front of the
detector cell.
9. Reconnect the signal and ignition cables to the
detector.
10. Reset the detector to the required operating
conditions. Refer to the relevant chapter of the
TM
TRACE
GC Ultra Operating manual for
more information.
Maintenance and Troubleshooting Manual
Chapter 11
Maintaining a FID
201