Replacement Procedure

GPIB Cable Replacement

GPIB Cable Replacement

Tools Required

Removal Procedure

Refer to
Step 1. Remove the 3/4 Rack Module Top as described in
Replacement" on page
Step 2. Disconnect the GPIB Cable connector (item 1) from PPM PCA card.
Step 3. Remove the two 7mm standoff (item 2) fastening the GPIB cable.
Step 4. Remove the GPIB Cable (item 3) from the Rear Panel Assembly.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
When you fasten the GPIB Cable with the two 7 mm standoff, set box
torque wrench's torque to 8.7 in-lb.
98
Torque screwdriver, TORX T20
Box torque wrench, 7 mm
Figure 6-13
for this procedure.
73.
"3/4 Rack Module Top
Keysight E4980A/E4980AL Precision LCR Meter

Hide quick links: