10 Alarms and error messages

10.2 Alarm messages

10.2.1 High-priority alarm (red)

Alarm
Airway pressure
Airway pressure
Apnea
Battery almost empty
Battery temperature critical
Device fault
Device temperature
196
EN
MEDUMAT Standard
Cause
Obstruction of the patient's
airways
Tube wrongly positioned
pMax set too low
Hoses kinked or pinched
Patient hose system leaking
Patient hose system not connected
correctly
Tube wrongly positioned
Hoses kinked or pinched
Ventilation settings incorrectly set Adjust ventilation settings.
Mask is not sitting correctly or is
leaking
No inspiration since the set apnea
alarm time, exception:
in Manual CPR mode (no
inspiration in the last 59 s)
Very low battery status
Battery temperature > 80°C
Temporary device malfunction
Device defective
Device temperature < -20°C
2
Remedy
Free the patient's airways of
obstructions.
Position tube correctly.
Adjust pMax.
Route hoses so that they are not
kinked or pinched.
Replace the patient hose system.
Connect patient hose system
correctly.
Position tube correctly.
Route hoses so that they are not
kinked or pinched.
Place the mask on tightly or
replace it.
Check the condition of the patient.
Select mandatory ventilation.
In CPR CCSV mode: Restart chest
compression.
Replace battery
(see 4.3.5,
Connect device to the line power
(see 4.2, p. 46)
and charge battery
(see 4.3.2,
p. 47).
Operate battery within the
permitted temperature range
(see 14.1.2,
p. 218).
Switch device off
(see 4.6,
p. 66)
and back on again
(see 4.5,
p. 65).
Perform a function check.
Have the device repaired.
Operate device within permitted
temperature range
(see 14.1.1,
p. 214).
p. 51).