Frequency Offset Operation (Option 089) - Agilent Technologies 8719ES User Manual

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

Operating Concepts
Knowing the Instrument Modes
Figure 7-46 Typical Test Setup for Tuned Receiver Mode
Tuned Receiver Mode In-Depth Description
If you press
, the analyzer receiver
System INSTRUMENT MODE
TUNED RECEIVER
operates independently of any signal source.
The following features and limitations apply to the tuned receiver mode:
• It is a fully synthesized receiver; it does not phase-lock to any source.
• It functions in all sweep types.
• It requires a synthesized CW source whose timebase is input to the analyzer's external
frequency reference.
For more information on using the tuned receiver mode, refer to
"Tuned Receiver Mode" on
page
2-27.

Frequency Offset Operation (Option 089)

Refer to
"Conversion Loss Using the Frequency Offset Mode" on page 2-13
for information
on frequency offset operation.
7- 85