Frequency Offset Operation - Agilent Technologies 8753ET User Manual

Network analyzers
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Operating Concepts
Knowing the Instrument Modes
Typical test setup
1. Activate the tuned receiver mode by pressing
TUNED RECEIVER
2. To perform a CW measurement using the tuned receiver mode, connect the equipment
as shown in
Figure
Figure 7-47 Typical Test Setup for Tuned Receiver Mode
Tuned Receiver Mode In-Depth Description
If you press
System INSTRUMENT MODE
operates independently of any signal source.
The following features and limitations apply to the tuned receiver mode:
• It is a fully synthesized receiver; it does not phase-lock to any source.
• It functions in all sweep types.
• It requires a synthesized CW source whose timebase is input to the analyzer's external
frequency reference.
For more information on using the tuned receiver mode, refer to
page
2-25.

Frequency Offset Operation

Refer to
"Conversion Loss Using the Frequency Offset Mode" on page 2-13
on frequency offset operation.
7-86
.
7-47.
INSTRUMENT MODE
System
TUNED RECEIVER
"Tuned Receiver Mode" on
, the analyzer receiver
for information