Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
7. After you connect the second ECal module, press
ECal.
8. Repeat steps 4 and 5 if you selected to calibrate using the manual thru option.
9. Review the displayed calibration results. Refer to
page
6-65.
10.Save the calibration results by pressing

Display the Module Information

Display information about the module by pressing
.
MODULE INFO
A text window is displayed that contains the following information about the selected
module.
1
PC interface unit port to which it is connected
2
Model number
3
Serial number
4
Connector type
5
Date of last certification
6-64
CONTINUE ECal
"Perform the Confidence Check" on
.
Save/Recall
ECal MENU
Cal
6
Number of measurement points
7
Start (minimum) frequency
8
Stop (maximum) frequency
9
Suggested warmup time
10
Warmup status (This is the remaining time
on an internal analyzer timer. The timer
simply counts down the recommended ECal
module warm-up time. When the timer
reaches zero, a message is displayed
indicating that the ECal module is ready. The
ECal module temperature or status is not
read. The timer is reset when the analyzer is
powered up or preset or when a new module
is attached.)