Removing the StepWave assembly from the source assembly

Required materials
Chemical-resistant, powder-free gloves
Seal breaker and locator tool
StepWave assembly removal and insertion tool
Recommendation:
the end of the StepWave assembly removal and insertion tool.
To remove the StepWave assembly from the source assembly:
Caution:
removing it from the source assembly, use only these tools:
• Seal breaker and locator tool
• StepWave assembly removal and insertion tool
Do not use any other tool.
1.
Position the seal breaker and locator tool so that it engages as follows:
Its pin with the hole on the adaptor housing directly under the end
of the StepWave ion guide assembly.
Its lip behind the StepWave ion guide assembly's ion guide cap.
4-54
Maintenance Procedures
When not in use, store the seal breaker and locator tool on
The source components can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
To avoid damaging the StepWave ion guide assembly when