ISE Default Values

Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
E
default: 0 mV
0
Low limit of E
: - 60 mV
0
High limit of E
: + 60 mV
0
Slope default: - 59.2 mV/decade
Slope range: - 62.0 mV/decade to - 45.0 mV/decade
C
: 35.5 ppm
ISO
Volume of flow cell: 95 mL
Concentration of standard 1 added during calibration = 2.96 %
Volume of standard 1 added during calibration = 0.50 mL
Concentration of standard 2 added during calibration = 2.96 %
Volume of standard 2 added during calibration = 1.50 mL