Table B-15. Default Analog Output Channel Assignments - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Thermo Fisher Scientific
set dout channel index state
This command assigns digital output channel (1-10) to be assigned to the
action associated with index (1-34), and assigns it an active state of state
(open or closed). Use the "list var dout" command to obtain the list of
supported index values and corresponding state. The following example sets
the digital output channel 4 to state 11.
Send:
set dout 4 11 open
Receive:
set dout 4 11 open ok
dtoa channel
This command reports the outputs of the 6 or 12 Digital to Analog
converters, according to
D/A #1 is 97.7% full-scale.
Send:
dtoa 1
Receive:
dtoa 1 97.7%
Note If the instrument is in a mode which does not provide a particular
output, and that output is selected, the value will be 0.0.
Note All channel ranges are user definable. If any customization has been
made to the analog output configuration, the default selections may not
apply.
Table B–15. Default Analog Output Channel Assignments
D to A
Function
1
Voltage Output
2
Voltage Output
3
Voltage Output
4
Voltage Output
5
Voltage Output
6
Voltage Output
7
Current Output
8
Current Output
9
Current Output
10
Current Output
11
Current Output
12
Current Output
Table
B–15. The following example shows that the
Single Range
Dual Range
CO
Low CO
CO
High CO
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
CO
Low CO
CO
High CO
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
C-Link Protocol Commands
I/O Configuration
Auto Range
CO
Range Status
Not Used
Not Used
Not Used
Not Used
CO
Range Status
Not Used
Not Used
Not Used
Not Used
Model 48i Instruction Manual
B-51