Use the following procedure to replace the kicker assembly
Equipment required:
Phillips drive, #1
1. Disconnect plumbing as shown.
2. Loosen 4-40 screw. Slide kicker assembly towards opening of keyway.
Equipment Damage Do not loosen or tighten this side of fitting as the
internal tubing will twist blocking flow. These fittings are covered by
silicone tape as a reminder.
3. Once kicker assembly clears the keyway, pull outwards towards the
board and up.
Figure 7–44. Removing and Replacing the Kicker Assembly
4. Replace the kicker assembly and assemble in reverse order.