The Pyrolyzer Reaction Tube; Replacement - Thermo Scientific SOLA II Flare User Manual

Sulfur online analyzer
Table of Contents

Advertisement

The Pyrolyzer
Reaction Tube

Replacement

Thermo Scientific
To replace the pyrolyzer tube (p/n 30-1025-0), first complete steps 1 through 7
in the previous section,
1. Gently lift the lid off the housing.
Note The pyrolyzer tube, insulation, and Kaowool insulation will be attached to
the lid. Carefully lift all of it out of the housing. ▲
Figure 8-15. Housing lid with insulation and pyrolyzer tube attached
2. To remove the existing pyrolyzer tube, unscrew the two fittings from
the lid while holding the tube in place. Remove the pyrolyzer tube and
set it in a safe location. Set the fittings, insulation, and Kaowool
insulation aside for use in the next steps.
3. Carefully push both ends of the replacement pyrolyzer tube through
holes of the insulation.
Figure 8-16. Pyrolyzer tube inserted into the insulation
Replacing the Pyrolyzer
Maintenance & Troubleshooting
Heater.
SOLA II Flare User Manual
Chapter 8
8-16

Advertisement

Table of Contents
loading

Table of Contents