9 Repair
9.5 Control panel
Easypad only
308
5. Update the nameplate at the control panel cover.
To do so, affix the supplied label as shown in the figure.
6. For "GALILEOS Comfort" (Easypad) only: Cable L10 (green cable)
must be equipped with ferrite core (B), unless this has already been
done.
9.5.1.1

What has to be done after replacing the user interface?

What has to be done after replacing the user interface?
Important: Cross-Reference GALILEOS
IMPORTANT: So that the board is also replaced with the user interface,
you MUST also follow the instructions in the chapter titled "Measures
following replacement of boards [ → 347]".
1. Check that the user interface and the display elements are
functioning correctly: When the unit is switched on, all of the display
elements must light up briefly.
2. Perform a software update to the latest version [ → 61].
GALILEOS with Easypad
Following replacement of the user interface, the language set on board
DX7 is set to the factory setting by default (00 = German, English, French,
Italian). If the configured unit language set (which can be queried by
running service routine S017.5 or using the "extended detail query" in
SiXABCon) has a configuration other than 00, this configuration will be
copied to board DX7 by the update function.
Sirona Dental Systems GmbH
Service Manual GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012