Laying of cables when replacing the ring motor
Laying of cables when replacing the ring motor
1. Lay the cable (G) parallel to cable L3 and secure it with the clamps.
2. Plug connectors X804 (I) and X813 (H) into board DX1.
9.3.4
What has to be done after replacing the ring motor
(M1_3)/pinion?
What has to be done after replacing the ring motor (M1_3)/pinion?
1. Check the function of the ring motor.
2. Perform complete unit adjustment or calibration [ → 155].