4339B

To Use Comparator Function

Setting the Limit Values

1. Press
2. Enter the lower limit value using the numeric ENTRY keys, then press
value. You can set the value from 9.900
3. Press
.
4. Enter the upper limit value using the numeric ENTRY keys, then press
value and to exit. You can set the value from 9.900

Sorting

To start sorting,
Press
. The Comprtr On annunciator(
To turn sorting OFF,
Press
again. The Comprtr On annunciator turns OFF.
The sorting results are HIGH, IN, and LOW.
Where,
HIGH
greater than higher limit
IN
between higher limit and lower limit
LOW
less than lower limit
The 4339B shows the comparison results using the display, beeper, printer, and
16064B LED Display/Trigger Box. (To use the 16064B, see \Accessories Available" later in this
chapter.)
For result output to the display, see \To Select Display Mode" in the next page.
For result output to the beeper, see \To Select Beeper Mode" in the next page.
For result output to the printer, see \To Print Measurement Data" later in this chapter.